Improvement of Calculating Medical Thermovision Camera Minimum Resolvable Temperature Difference

Authors

  • Ахмед Малік Лазім Аль-Мзіраві NTUU KPI, Ukraine
  • Валентин Георгійович Колобродов NTUU KPI, Ukraine
  • Володимир Іванович Микитенко NTUU KPI, Ukraine

DOI:

https://doi.org/10.20535/1810-0546.2014.5.35252

Keywords:

Medical thermal imager, Minimum resolvable temperature difference, Modulation transfer function

Abstract

The paper is devoted to improvement of the equation for calculating the minimum resolvable temperature difference (MRTD) of medical thermovision camera by means of more sophisticated models of thermal images visual perception. An MRTD calculating algorithm, which is based on a more reliable approximation of the visual system modulation transfer function by Schultz, is considered. We obtain new expression for the bandwidth of Schultz’s approximation, which does not depend on the angular size of the Foucault pattern bar. An improved equation for calculating the MRTD was obtained. Analysis of the equation has showed that it coincides with the well-known equations for Foucault pattern bar with appointed angular size. The limits of validity of this equation in the visual system spatial integration from 1 to 5,7 arc. minute were determined. An example calculation for Thermal Eye TSC Imager MRTD was fulfilled. There were received two differing analytical MRTD equations, which use Schultz and Lloyd approximations. It is shown that these equations coincide with each other when pixel angular size of microbolometer matrix is equal to 0,71 mrad.

Author Biographies

Ахмед Малік Лазім Аль-Мзіраві, NTUU KPI

Postgraduate student 

Валентин Георгійович Колобродов, NTUU KPI

Doctor of technical sciences, professor, head of department of the NTUU KPI

Володимир Іванович Микитенко, NTUU KPI

Candidate of sciences (engineering), associate professor, senior research fellow, assistant professor at the NTUU KPI

References

G.C. Wishart et al., “The accuracy of digital infrared im-aging for breast cancer detection in women undergoing breast biopsy”, European J. Surgical Oncology, vol. 36, no. 6, pp. 535–540, 2010.

R.H. Vollmerhausen et al., Analysis and evaluation of sampled imaging systems. SPIE Press, 2010, 304 p.

Ллойд Дж. Системы тепловидения / Пер. с англ. – М.: Мир, 1978. – 416 с.

Колобродов В.Г., Лихоліт М.І. Проектування тепловізійних і телевізійних систем спостереження: Підручник. – К.: НТУУ “КПІ”, 2007. – 364 с.

Мирошников М.М. Теоретические основы оптико-электронных приборов: Учеб. пособие для приборостроительных вузов. – 2-е изд., перераб. и доп. – Л.: Машиностроение, 1983. – 696 с.

Тарасов М.М., Якушенков Ю.Г. Инфракрасные системы "смотрящего" типа. – М.: Логос, 2004. – 444 с.

Published

2014-11-19

Issue

Section

Art